Clinical experiments

Measurement of microbubbles in the return line after the venous chamber
verified high amounts of microbubbles that were higher/worse if the blood pump
speed used for the specific patient was higher. (Stegmayr 2007)

Since it was generally stated that air bubbles resorbed when entering the body we
performed a study microbubble count measuring bubble count by a hand-held
probe and ultrasound signals. A significant increase in microembolic signals were
found after start of dialysis at the vascular access but also at the external carotic
artery (Forsberg 2010)

In a clinical interventional study we showed that using a wet dialyzer caused less
microbubble contamination in the blood return line of the patient. In addition the
use of a low blood level in the venous chamber of the Fresenius 5008 blood line
cause significantly more microbubble counts in the return line (Forsberg 2013).

A follow up and blinded study using a high versus low blood level in the venous
chamber showed significant better results for a high blood level (Forsberg 2012).

In an overview study ultrasound measurements were performed to
evaluate where microbubble counts were expected to enter the
extracorporeal bloodline during dialysis showed addition of signals at
the vascular access site, within the bloodline where a negative
pressure existed, after the dialyzator, after the venous chamber
before blood entry into the bloodline of the patient (Jonsson, Drybcak
2022).

In another study foam was detected in the Artis bloodline during
dialysis. Follow up studies revealed a problem with the autopriming
and advices to the manufacturer who updated the autopriming
software. The extent of visible foam was greater with a larger surface
area of the dialyzer and for HDF dialyzers(Jonsson 2018).

The Swedish Medical Product agency approved a clinical study using
the Emboless® venous chamber and to compare with the Fresenius
5008 bloodline venous chamber. The randomized self-cross over
study revealed a significantly better removal of microbubbles at the
outlet line of the venous chamber with Emboless ( a median of 76%
reduction) versus with the Fresenius 5008 chamber ( a median of
39%, p<0.001).