Microbubble complications

Air contamination of the blood of the patient appears in conjunction with infusion of fluid, injections of drugs
and during extracorporeal treatments such as dialysis, apheresis, and cardio-pulmonary bypass.

Larger amounts may lead to instant death while smaller amounts such as microbubbles will cause microemboli
by the gas bubble itself and further by the attached clot that appears around and attached to the bubble (See
figure).

A single microemboli will cause a local damage in the tissue where it deposits. If this is the lung a scar will
develop but experienced side effects may recover with no or limited side effects to the patient. If the
localization of the microemboli is in a crucial part of the body such as regulating center in the heart or brain
severe consequences may appear, such as sudden death or a crucial stroke.

If numerous microbubble exposures appears tissue damage will sum up with time causing progressive organ
damage.

To limit gas exposure the concept of including a venous chamber in a blood line before the blood or infusion
fluid returns to the patient was introduced to enable air/gas to evaporate out of the fluid already in the
1950th…?

However, we have investigated venous chambers in clinical use for dialysis and found that the efficacy to
remove air bubbles is strongly limited (See laboratory work, Jonsson et al 2023). Most dialysis devices do not
alarm for microbubble contamination if the size of a bubble is less than 500µm diameter.