Laboratory experiments

The contamination of air within the closed extracorporeal dialysis system was described Jonsson in his Thesis. During – recirculation of dialysis priming fluid ? in … minutes 6 ml of air was consolidated and isolated from the circuit.

A viscous fluid was prepared by saline, sepharos …. and albumin ..%, to be used at room temperature but to mimic the viscosity of the blood (referens). This fluid was able to show that ultrasound detection of microemboli were due to microbubbles (the fluid is devoid of blood and therefore counts couldn’t be made by blood clots).

It was now possible to visualize larger microbubbles passing the air trap without inducing an alarm and entering the sieve of the venous chamber. (see movie).

Follow up experiments visualized that the shape of air traps and the pump speed of the fluid in the extracorporeal circuit influenced the contamination of microbubbles of air at the Inlet and Outlet side (C Stegmayr et al 2007).

Further a primed but originally dry dialyzer caused a greater amount of inlet microbubbles than a wet stored gamma sterilized dialyzer (C Stegmayr et al 2008).

Priming of more than 3 L was necessary to achieve a considerable reduction of bubbles in the extracorporeal circuit (C Stegmayr et al )

These results showed that air contamination appeared into the blood by a whipping cream effect caused by the fluid entering the vanous chamber from the top (see figure)

Various experimental studies were performed to define differences in shape to optimize reduction of microbubbles by a specific chamber for the venous blood return. This resulted in a designs that was patented and named Emboless® (oklart hur mycket vi refererar till).

A follow up in vitro study was performed to compare four frequently used extracorporeal circuits, with different shapes of venous chambers, with the Emboless® chamber. Compared devices and their extracorporeal tubing systems were Baxter Artis, Baxter AK78, Fresenius 5008 and Fresenius 6008.

The Emboless most efficiently eliminated the microbubbles (Jonsson 2023). Again a higher blood pump speed had significant worse impact on the outcome.